Configure and execute a 10-step linear list
sweep
The following example demonstrates the steps to set up the power
supply to output a linear list sweep from 1 V to 10 V in 1 V steps
(default settings). Each step in the list sweep will dwell for 3 seconds.
The list will be stored in list location 1. The output will turn on at 0 V
and return to 0 V at the end of the list execution.
To configure a list on the front panel:
1. Press the MENU key.
2. Under Measure, highlight the L ist icon using either the
navigation control or the soft keys. Press ENTER.
3. Select point 1 and press ENTER to put a check mark in the
box.
4. Highlight Insert and press ENTER to increment the points
display until it reads 10.
5. Select the Voltage box in the point 1 row. Press ENTER.
6. Make the value 1 V and press ENTER.
7. Select the Current box in the point 1 row. Press ENTER.
8. Make the value 2 A and press ENTER.
9. Select the Dwell box in the point 1 row. Press ENTER.
10. Make the value 3 s and press ENTER.
11. Repeat steps 5 to 8 for points 2 to 10 and increase the voltage
by 1 V at each higher step.
12. Select the Save button to store the list in memory location 1.
13. Enter a name for the list on the displayed keypad and press
ENTER.
14. Press the EXIT key.
To delete a row on the front panel:
1. Highlight the Points cell of the row you want to delete.
2. Press ENTER to select the cell row.
3. Navigate to Delete.
4. Press ENTER.