2600AS-900-01 Rev. B / September 2008 Return to Section Topics 3-13
Series 2600A System SourceMeter
®
Instruments User’s Manual Section 3: Test Script Processor Interaction
The Series 2600A command sequence:
Figure 3-6
LabVIEW source step example
localnode.prompts = 0 -- Disable prompts.
data_table = {} -- Clear table.
reset() -- Reset Series 2600A.
smua.source.output = 1 -- Turn on source output.
for i=1,10 do -- Loop for 10 steps.
smua.source.levelv = i -- Set source level.
mi = smua.measure.i() -- Take a current measurement.
table.insert(data_table,mi) -- Store measurement in table.
end -- End of loop.
smua.source.output = 0 -- Turn off output.
print(table.concat(data_table,’,’)) -- Print table of current values.
reset() -- Reset instrument.