Power and environmental specications 
For indoor use only.
Power Supply 100 VAC to 240 VAC, 50 Hz to 60 Hz 
(auto sensing)
Maximum VA 240 VA
Operating altitude Maximum 2000 m (6562 feet) above 
sea level
Operating 
temperature
0 °C to 50 °C (32 °F to 122 °F),  
70% relative humidity up to 35 °C. 
Derate 3% relative humidity, 
35 °C to 50 °C
Storage temperature 25 °C to 65 °C (77 °F to 149 °F)
Pollution degree 1 or 2
DC source output 30.3 W maximum
  ±200 VDC maximum
  ±1.515 A at ±20 VDC
  ±101 mA at ±200 VDC
Pulse region 4* Region maximums: 10 A at 5 V
Maximum pulse width: 1 ms
Maximum duty cycle: 2.2%
*See full specications for other pulse regions.
Carefully consider and congure the appropriate 
output-off state, source levels, and compliance 
levels before connecting the instrument to 
a device that can deliver energy. Failure to 
consider the output-off state, source levels, and 
compliance levels may result in damage to the 
instrument or to the device under test. 
Carefully consider and congure the appropriate 
output-off state, source levels, and compliance 
levels before connecting the instrument to 
a device that can deliver energy. Failure to 
consider the output-off state, source levels, and 
compliance levels may result in damage to the 
instrument or to the device under test. 
Measurement input Measurement category I
Voltage: 200 VDC maximum HI to LO, 
250 VDC LO to ground
Current: 101 mA at 200 V; 
1.515 A at 20 V
Impedance: Variable