Series 3700A System Switch/Multimeter Reference Manual Section 4: Basic DMM operation
3700AS-901-01 Rev. D/June 2018 4-23
• DMM limit low value (–4294967295 to +4294967295)
• line synchronization (dmm.ON or dmm.OFF)
• math enable (dmm.ON or dmm.OFF)
• math format (dmm.MATH_NONE, dmm.MATH_MXB, dmm.MATH_PERCENT, or
dmm.MATH_RECIPROCAL)
• math mxb b factor (–4294967295 to +4294967295)
• math mxb m factor (–4294967295 to +4294967295)
• math mxb units
• math percent (–4294967295 to +4294967295)
• nplc (0.0005 to 15 for 60 Hz; 0.0005 to 12 for 50 Hz)
• offset compensation (dmm.ON or dmm.OFF)
• open detector (dmm.ON or dmm.OFF)
• reference junction (dmm.REF_JUNCTION_SIMULATED, dmm.REF_JUNCTION_INTERNAL, or
dmm.REF_JUNCTION_EXTERNAL) (only available when transducer type is set to
thermocouple).
• relative offset enable (dmm.ON or dmm.OFF)
• relative offset level
• RTD alpha (0 to 0.01) (only with dmm.transducer set to dmm.TEMP_FOURRTD or
dmm.TEMP_THREERTD)
• RTD beta (0 to 1.0) (only with dmm.transducer set to dmm.TEMP_FOURRTD or
dmm.TEMP_THREERTD)
• RTD delta (0 to 5) (only with dmm.transducer set to dmm.TEMP_FOURRTD or
dmm.TEMP_THREERTD)
• RTD zero (0 to 10000) (only with dmm.transducer set to dmm.TEMP_FOURRTD or
dmm.TEMP_THREERTD)
• simulated reference temperature (Celsius (0 °C to 65 °C), Fahrenheit (32 °F to 149 °F), or Kelvin
(273 °K to 338 °K)) (only with dmm.transducer set to dmm.TEMP_THERMOCOUPLE)
• thermistor type (2252, 5000, or 10000) (only with dmm.transducer set to
dmm.TEMP_THERMISTOR)
• thermocouple type (dmm.THERMOCOUPLE_J, dmm.THERMOCOUPLE_K, dmm.THERMOCOUPLE_T,
dmm.THERMOCOUPLE_E, dmm.THERMOCOUPLE_R, dmm.THERMOCOUPLE_S,
dmm.THERMOCOUPLE_B, dmm.THERMOCOUPLE_N) (only with dmm.transducer set to
dmm.TEMP_THERMOCOUPLE)
• three-wire RTD type (dmm.RTD_PT100, dmm.RTD_D100, dmm.RTD_F100, dmm.RTD_PT385,
dmm.RTD_PT3916, dmm.RTD_USER) (only with dmm.transducer set to dmm.TEMP_THREERTD)
• transducer type (dmm.TEMP_THERMOCOUPLE, dmm.TEMP_THERMISTOR, dmm.TEMP_THREERTD,
or dmm.TEMP_FOURRTD)
• units (dmm.UNITS_CELSIUS, dmm.UNITS_KELVIN, or dmm.UNITS_FAHRENHEIT)