Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-43
6. Send reading.
7. Set voltage, –12 V to +12 V range, generate positive point 2.
8. Send reading.
9. Set current, 0 mA to +20 mA range, generate point 1.
10. Send reading.
11. Set current, 0 mA to +20 mA range, generate point 2.
12. Send reading.
13. Set current, +4 mA to +20 mA range, generate point 1.
14. Send reading.
15. Set current, +4 mA to +20 mA range, generate point 2.
16. Send reading.
For totalizer channels, a calibration sequence includes these steps:
1. Calibrate 0 V totalizer threshold
2. Calibrate 1.5 V totalizer threshold
You must save the calibration after calibrating and before locking. Use
channel.calibration.save() to save the calibration.
All calibration progress is lost if the calibration data is not saved before you lock the channel.
After calibration, the channel must be locked using channel.calibration.lock().
Also see
channel.calibration.lock() (on page 11-40)
channel.calibration.save() (on page 11-41)
channel.calibration.unlock() (on page 11-44)

Table of Contents

Other manuals for Keithley 3700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3700 Series and is the answer not in the manual?

Keithley 3700 Series Specifications

General IconGeneral
BrandKeithley
Model3700 Series
CategoryMultimeter
LanguageEnglish

Related product manuals