SrcA:+00.0000 V LimA:100.000mA
AUTO
SrcA:+20.0000 V LimA:100.000mA
AUTO
Source value set to 20.0000 V
3. Press the CURSOR keys to move the cursor to the
10s digit.
4. Press the navigation wheel
P
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
P
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
to enter EDIT mode.
The EDIT indicator appears in the upper left corner
of the display.
5. Turn the navigation wheel
P
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
P
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
to set the source value to
20.0000 V, and then press the navigation wheel
P
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
P
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
to enter
the selection and exit EDIT mode.
The main display screen reappears: Step 2: Set the source limit
1. Press the LIMIT
key. You will see a blinking character in
the LimA value eld.
2. While that character is blinking, press the RANGE
keys
as needed to select the 10 mA limit range. Verify that the
source limit value in the LimA eld is 10.0000 mA.
SrcA:+20.0000 V LimA:10.0000mA
AUTO
Source limit set to 10.0000mA

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2600B Series and is the answer not in the manual?

Keithley SourceMeter 2600B Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals