Carefully consider and congure the appropriate output-
o󰀨 state, source levels, and compliance levels before
connecting the instrument to a device that can deliver
energy. Failure to consider the output-o󰀨 state, source
levels, and compliance levels may result in damage to the
instrument or to the device under test.
Power supply
100 VAC to 240 VAC, 50 Hz to 60 Hz
(autosensing)
Maximum VA
240 VA
Operating altitude
Maximum 2000 m (6562 ft) above
sea level
Operating
temperature
0 °C to 50 °C, 70% relative humidity
up to 35 °C
Derate 3% relative humidity, 35 °C to 50 °C
Storage temperature
25 °C to 65 °C
Pollution degree
1 or 2
DC source output
30.3 W maximum
±200 VDC maximum
±1.515 A at ±20 VDC
±101 mA at ±200 VDC
Pulse region 4*
Region maximums: 10 A at 5 V
Maximum pulse width: 1 ms
Maximum duty cycle: 2.2%
Power and environmental specications
For indoor use only.
Measurement
input
Measurement category O
Voltage: 200 VDC maximum HI to LO,
250 VDC LO to ground
Current: 101 mA at 200 V;
1.515 A at 20 V
Impedance: Variable
*See full specications for other pulse regions

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2600B Series and is the answer not in the manual?

Keithley SourceMeter 2600B Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals