EasyManuals Logo

Keithley SourceMeter 2602 Reference Manual

Keithley SourceMeter 2602
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #337 background imageLoading...
Page #337 background image
Series 2600 System SourceMeters Reference Manual Instrument Control Library 12-63
Return to Section 12 topics 2600S-901-01 Rev. A / May 2006
Example Selects positive calibration constants for all subsequent measurements:
smua.cal.polarity = smua.CAL_POSITIVE
smuX.cal.restore
X = SMU channel (a or b)
Function Loads a stored set of calibration constants.
Usage There are two ways to use this function:
smuX.cal.restore()
smuX.cal.restore(calset)
calset Calibration set to be loaded.
Set calset to one of the following values:
smuX.CALSET_NOMINAL A set of calibration constants that are uncalibrated but
set to nominal values to allow rudimentary functioning
of the instrument.
smuX.CALSET_FACTORY The calibration constants when the instrument left the
factory.
smuX.CALSET_DEFAULT The normal calibration set.
smuX.CALSET_PREVIOUS The calibration set that was used before the last default
set was overwritten.
If calset is not specified, smuX.CALSET_DEFAULT will be used.
Remarks
• This function will overwrite the current set of calibration constants from non-volatile
memory.
• This function will be disabled until a successful call to smuX.cal.unlock is made.
Details See Section 16 (calibration).
Also see smuX.makebuffer
Example
Restores factory calibration for SMU A:
smua.cal.restore(smua.CALSET_FACTORY)

Table of Contents

Other manuals for Keithley SourceMeter 2602

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2602 and is the answer not in the manual?

Keithley SourceMeter 2602 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2602
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals