470-BA-defs-0017
38
The indication of a weight unit, i.e. „g“ can be changed to another unit, e.g. 2kg“ by
repeatedly pressing
F
.
Setting unit see point 10 „General programmations“:
Weighing with tolerance range, see point 7.
Weighing with accumulation, see point 8.
6.1.2 Taring TARE
Place tare cup on the weighing plate and press
TARE
.
The weighing indication changes now to
0.
The weight of the cup is now memorised.
Fill weighing object in the cup, read measuring value.
By pressing
TARE
, after the terminated weighing operation
0
appears again in
the display.
The taring procedure can be repeated several times, for instance to produce a mixture
of several components.
The limit is reached when the total weighing range has been used.
After removal of the tare cup the total weight appears as a negative value with „-„
mark.
6.1.3 Bar graph display when weighing
Will be activate with weight unit „B“ is active on the display. This is available by F. See
point 6.1.1 „Weighing“.
The bar graph (below the digital display) moves from the left to the right side and
mar5ches in the same measure as the balance is being loaded. Is full breadth will be
reached at full capacity.
This is to display analogously the actual remaining weighing range and to avoid an
overloading.
6.1.4 Net-total-weighings
Useful when a mixture of several components will be weighed in a tare cup, and when
finally the sum weight of all components will be needed for control (Net-total), i.e.
without the weight of the tare cup.
See point 8 „Accumulation of indicated values“, point 8.2 „Net-total-weighings“