9.2 Numeric entering of the reference weight
If you know the reference weight/piece you can enter this via number keys.
Enter reference weight via number keys
0�
a
PRE-
TARE
1.833
0
M+
Confirm with key
0�
a
PRE-
TARE
1.833
Reference weight
0
M+
Now you can place the parts to be counted onto the weighing plate. All quantity pa-
rameters of your goods to be weighed are displayed:
9.3 Automatic reference optimization
If it was impossible to determine a reference due to instable goods to be weighed or
an insufficient reference weight, the [
] display will appear in the reference weight
window during reference calculation.
Overlay
indicates:
Placed number of pieces insufficient for reference calculation
CXB models < 40 d
CXP models < 20 d
Placed reference weight insufficient for reference calculation
CXB models < 4/5 d
CXP models < 1/5 d
Add additional parts until the [
] display disappears.
An audio signal indicates that reference optimization has been carried out.
At every reference optimisation, the reference weight is calculated anew. As the ad-
ditional pieces increase the base for the calculation, the reference also becomes
more exact.