- 10 - EOB-B_EOE-B_EOS-B-SH-e-1911
The specified adjustment weight is displayed flashing.
(example)
Place the adjustment weight centrically on the weighing pan and
confirm with . "CAL Y" will flash briefly and a signalling
tone will sound. Adjustment will be performed. Then the balance
returns automatically into the weighing mode.
An error message will be displayed in the event of an adjustment error or incorrect
adjustment weight. Remove the adjustment weight and repeat the adjustment
process.
Keep the adjustment weight close to the balance. Daily control of the weighing
exactness is recommended for quality-relevant applications.
Operation
Start-up
Start balance by pressing .
The balance will carry out a self-test As soon as the weight
display appears, the balance is ready for weighing.
Should the balance not display exactly zero despite empty weighing pan, press the
button. The balance will be set to zero after a short standby time.
Switching Off
Switch off balance with , the display will go off.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN EOB Series and is the answer not in the manual?

KERN EOB Series Specifications

General IconGeneral
Weighing plate materialstainless steel
Power supply/adapter includedyes
Protection ClassIP65
CalibrationExternal calibration
DisplayLCD
InterfaceRS-232
Rechargeable batteryYes