102
770/GS/GJ-BA-defsi-0023
4.3 Weighing in Percent
Weighing in Percent Code 2 1 5
Symbol displayed:
%
This application program enables you to obtain weight readouts in percent which are in
proportion to a reference weight. The reference weight is stored as a menu-defined
percentage (factory setting: 100 %).
Changing the Reference Percentage
The reference percentage can be changed by you, in cycles. Choose from the following
settings: 5, 10, 20, 50 and 100.
To activate this function:
Hold down the
F
key for more than 2 seconds until “
rEF
100%
“ appears on the display.
To change the setting:
Briefly press the
F
key.
To store the percentage:
Hold down the
F
key for more than 2 seconds.
When you exit the reference storage mode, “
Err 22
“ may briefly be displayed. This
means that the new reference value has been stored.
This setting is
not
cancelled by the reset function 9--1
o
!
Display Parameter for Readouts in Percent
The following display parameters can be set for readouts in percent:
The readout in percent is displayed... Code
without a decimal place 3 6 1
with one decimal place 3 6 2*
with two decimal places 3 6 3
with three decimal places 3 6 4
The number of decimal places is automatically decreased, if the weight stored is too
light to be displayed.
* = factory setting