CKE-BA-e-2132 41
15 Help for any minor failures
If there are any programme execution problems, the scales should be switched off
and disconnected from the mains for a while. Next, the weighing process should be
started anew.
Problem Possible cause
The weight indicator
is not lit
The scales is not on.
Interrupted mains connection (mains cable not
connected/damaged).
Mains voltage failure.
The weight indication
keeps fluctuating
Draft/air movements.
Table/air vibrations.
The scales plate is in contact with foreign bodies.
Electromagnetic fields / static discharge (select
another location / if possible, switch off the interfer-
ing device).
The weighing result
is clearly wrong
The scales indication was not reset.
Incorrect adjustment.
Scales not placed on a level surface.
There are heavy temperature fluctuations.
The heating time not observed.
Electromagnetic fields / static discharge (select
another location / if possible, switch off the interfer-
ing device).

Table of Contents

Other manuals for KERN TCKE 300-3-A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN TCKE 300-3-A and is the answer not in the manual?

KERN TCKE 300-3-A Specifications

General IconGeneral
BrandKERN
ModelTCKE 300-3-A
CategoryScales
LanguageEnglish

Related product manuals