4 - Language Dictionary
44
Calibration Commands
Calibration commands let you:
Enable and disable the calibration mode
Change the calibration password
Calibrate the current and voltage programming and measurement, and store new calibration
constants in nonvolatile memory.
NOTE: If calibration mode has not been enabled with CALibrate:STATe, programming the
calibration commands will generate an error.
CALibrate:CURRent
This command initiates the calibration of the positive dc output current as well as the high-range current
measurement circuit.
Command Syntax
CALibrate:CURRent[:SOURce][:DC][:POSitive]
Parameters
None
Examples
CAL:CURRCAL:CURR:SOUR:DC:POS
Related Commands
CAL:CURR:NEG
CALibrate:CURRent:NEGative
This command initiates the calibration of the negative dc output current.
Command Syntax
CALibrate:CURRent[:SOURce][:DC]:NEGative
Parameters
None
Examples
CAL:CURR:NEGCAL:CURR:SOUR:DC:NEG
Related Commands
CAL:CURR
CALibrate:CURRent:MEASure:LOWRange
This command initiates the calibration of the low-range current measurement circuit.
Command Syntax
CALibrate:CURRent:MEASure[:DC]:LOWRange
Parameters
None
Examples
CAL:CURR:MEAS
Related Commands
CAL:CURR
CALibrate:CURRent:MEASure:AC
Keysight 66312A, 66332A Only
This command initiates the calibration of the high bandwidth (ac) measurement circuit.
Command Syntax
CALibrate:CURRent:MEASure:AC
Parameters
None
Examples
CAL:CURR:MEAS:AC