Language Dictionary - 4
SENSe:FUNCtion
Keysight 66312A, 66332A Only
This command configures the measurement sensor to measure either voltage or current when an acquire
trigger is used. The query returns the function setting, either VOLT or CURR.
SENSe:SWEep:OFFSet:POINts
Keysight 66312A, 66332A Only
This command defines the offset in a data sweep when an acquire trigger is used. Negative values
represent data samples taken prior to the trigger.
SENSe:SWEep:OFFSet:POINts <NRf+>
-4095 through 2,000,000,000
SENSe:SWEep:OFFSet:POINts?
SENS:SWE:TINT SENS:SWE:POIN MEAS:ARR
SENSe:SWEep:POINts
This command defines the number of points in a data sweep.
SENS:SWE:TINT SENS:SWE:OFFS MEAS:ARR
SENSe:SWEep:TINTerval
This command defines the time period between samples. The value that you enter for the time interval will
be rounded to the nearest 15.6 microsecond increment.
SENSe:SWEep:TINTerval<NRf+>
15.6 microseconds through 31200 seconds
SENS:SWE:POIN SENS:SWE:OFFS MEAS:ARR