4 - Language Dictionary
CALibrate:STATe
This command enables and disables calibration mode. The calibration mode must be enabled before the
will accept any other calibration commands.
The first parameter specifies the enabled or disabled state. The second parameter is the password. It is
required if the calibration mode is being enabled and the existing password is not 0. If the password is not
entered or is incorrect, an error is generated and the calibration mode remains disabled. The query
statement returns only the state, not the password.
NOTE: Whenever the calibration state is changed from enabled to disabled, any new calibration
constants are lost unless they have been stored with CALibrate:SAVE.
CALibrate:STATe<bool>[,<NRf>]
0 | 1 | OFF | ON [,<password>]
CAL:STAT 1,6812 CAL:STAT OFF
CALibrate:VOLTage
This command initiates the calibration of the output voltage and the voltage readback circuit.
CALibrate:VOLTage:PROTection
This command can calibrates the overvoltage protection (OV) circuit. The dc source automatically
performs the calibration. CALibrate:VOLTage:PROTection is a sequential command that takes several
seconds to complete.
CALibrate:VOLTage:PROTection