Installation Note N9010-90067 3
Option EXM, External Mixing Upgrade Kit
Option EXM, External Mixing Upgrade Kit
Software and test equipment is required for making adjustments and for performance verification
testing.
Information on how to obtain this software can be found at:
www.keysight.com/find/calibrationsoftware
While Keysight does recommend that a full calibration be performed after the installation of this
upgrade, the end user must ultimately determine whether they want this service or not. If a full
calibration is required, arrangements regarding the level of calibration must be made between the
end user and the calibration provider.
Introduction
This kit includes parts to upgrade an EXA signal analyzer with frequency range Option 532 or 544 to
add Option EXM, External Mixing.
In addition to installing the hardware to support Option EXM, a license for the following option will
also be installed:
— N9010B-EXM, External Mixing
Products Affected: N9010B EXA
Serial Numbers: All
To Be Performed By: (X) Keysight Service Center
(X) Personnel Qualified by Keysight
( ) Customer
Estimated Installation Time:
Estimated Adjustment Time:
Estimated Verification Time:
1.5 Hours
1.0 Hours
a
1.0 Hours
a
a. To ensure that these newly installed options are functioning properly, the procedure that follows
includes the requirement of performing certain adjustments and performance verification tests.
However, the completion of these tests does not guarantee that the instrument meets all advertised
specifications.