Installation Note N5222-90131 85
Description of the Upgrade
Installation Procedure for the Upgrade
Navigate to the EEPROM Backup Utility, located at:
Windows 7 -- C:\Program Files (x86)\Keysight\Network
Analyzer\Service\eebackup.exe
Windows 10 -- C:\Program Files\Keysight\Network
Analyzer\Service\eebackup.exe
Run the program.
Click Backup EEPROM.
Click Backup TSMB Mem.
Click Exit when the program has finished.
Operator’s Check
Perform the Operator’s Check to check the basic functionality of the analyzer.
For instructions, click the Chapter 3 bookmark “Tests and Adjustments” in the
PDF Service Guide
1
.
If you experience difficulty with the basic functioning of the analyzer, contact
Keysight. Refer to “Contacting Keysight” on page 4.
Calibration
Although the analyzer functions, its performance relative to its specifications
has not been verified. It is recommended that a full instrument calibration be
performed using the analyzer’s internal performance test software. To view
information on the performance test software, click the Chapter 3 bookmark
“Tests and Adjustments” in the PDF Service Guide
1
.