8 Chapter 1
Functional Tests
Test Equipment
Preamp
Options
(Pxx)
30 dB Fixed Attenuator Accuracy: < 0.05 dB @ 50 MHz 11708A
Cables
All 3.5 mm (1 meter) Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
Loss: < 2.0 dB
11500E 8120-4921
543, 544,
550
2.4 mm (f) to 2.4 mm (m)
(1 meter)
Frequency: 10 MHz to 50 GHz
VSWR: < 1.55:1 at 50 GHz
Loss: < 6.0 dB
8120-6164
BBA Cable, BNC BNC male
Length: 23 cm (9 in) max
10502A
All Cable, BNC
(3 required)
120 cm (48 in.) BNC cable 10503A
Signal Source
All Synthesized Sweeper
Frequency: 10 MHz to 50 GHz
a
Harmonic level: < − 40 dBc
Amplitude range: 10 to − 20 dBm
Frequency Accuracy: 0.02%
PSG
b
83630B,
83640B,
83650B
BBA RF Signal Generator Frequency: 250 kHz to 40 MHz
Amplitude: ~–10 dBm
Flatness: < ±0.75 dB
E4438C 33250A
Power Meter
All Power Meter Power Reference Accuracy: ±1.2%
Compatible with power sensor
E4418B E4419B
All Power Sensor Frequency Range: 50 MHz to 3.66 GHz
Amplitude Range: −70 to –20 dBm
8481D 8487D
8485D
503, 508,
513
Power Sensor Frequency Range: 50 MHz to
13.66 GHz
Amplitude Range: −30 to +20 dBm
8481A
526 Power Sensor Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: −30 to +20 dBm
8485A
543, 544,
550
Power Sensor Frequency Range: 50 MHz to 50 GHz
Amplitude Range: −30 to +20 dBm
8487A
P03 Low Power Sensor Frequency Range: 50 MHz to 3.6 GHz
Amplitude Range: –70 to –20 dBm
8481D 8485D
8487D
Table 1-1
Analyzer
Option
Item Critical Specifications
Recommended
Keysight Model
Alternate
Keysight
Model