EasyManua.ls Logo

Keysight Technologies X Series - Page 48

Keysight Technologies X Series
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44 Chapter 8
Scale Fidelity
Procedure
Table 8-1 Scale Fidelity Results
External Attenuator Setting
Minimum
(dB)
Marker Delta Value
(dB)
Maximum
(dB)
0 N/A Reference N/A
10 11.0 9.0
20 21.0 19.0
30 31.0 29.0
40 41.0 39.0
50 51.0 49.0