76 Keysight CXG, EXG, and MXG X-Series Signal Generators Programming Guide
Programming Examples
GPIB Programming Interface Examples
ViStatus viStatus = 0;
// Opens a session to the GPIB
device
// at address 19
viStatus=viOpenDefaultRM(&defaultRM);
viStatus=viOpen(defaultRM, "GPIB::19::INSTR", VI_NULL, VI_NULL,
&vi);
if(viStatus){
printf("Could not open ViSession!\n");
printf("Check instruments and connections\n");
printf("\n");
exit(0);}
viPrintf(vi, "*RST\n"); // initializes signal generator
// prints to the output window
printf("The signal generator should now be in REMOTE. The remote
indicator\n");
printf("annunciator R should appear on the signal generator
display\n");
printf("\n");
viClose(vi); // closes session
viClose(defaultRM); // closes default session
}
Local Lockout Using HP Basic and GPIB
This example demonstrates the Local Lockout function. Local Lockout disables
the front panel signal generator keys. basicex2.txt performs the following
functions:
— resets instrument
— places signal generator into local
— places signal generator into remote
The following program example is available on the signal generator
Documentation CD-ROM as basicex2.txt.
10
!******************************************************************
20 !

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals