Electronic Calibration of the 3458A (Product Note 3458A-3) B
Keysight 3458A Calibration Manual 121
Offset adjustments
To remove offset errors of internal circuits, the 
multimeter internally connects a short in place of 
the input terminals and measures the offset. 
Normal measurements of signals on the input 
terminals subtract this offset to give the correct 
reading. The only offset errors not removed by 
this approach are thermocouple offsets along the 
path from the input terminals to the point that is 
switched to the internal short.
These errors require a four-wire short on both the 
front and rear input terminals (switch selected). 
With these external inputs, one command, CAL 0, 
executes zero offset measurements that result in 
additional offset calibration constants to correct 
subsequent readings.
Other multimeters use this approach to removing 
offset errors. The 3458A multimeter simply makes 
improvements by using more stable components, 
again minimizing time and environmental errors.
DC gain adjustments
Gain adjustments of all five DC voltage ranges 
(100 mV full- scale to 1000 V full-scale) require 
only one external DC voltage standard. The DC 
voltage input path, shown in Figure 2, requires 
three adjustments, potentially. The product, 
represents the calibration gain used on any given 
range.
Internal tolerance limits for each gain adjustment 
are factory set. A gain value outside the 
associated tolerance indicates a malfunctioning 
instrument. Therefore, as the gain adjustments 
are being computed, the instrument checks the 
value of each gain adjustment and flags errors 
accordingly.
Figure 2.
The DC input path to the analog-to-digital converter is either 
through an amplifier only or through an additional resistive 
attenuator, depending on the range used.
The user enters the exact value of the external  
10 V DC voltage standard (for example, “CAL 
10"). The following sequence, performed 
automatically by the 3458A multimeter, 
determines gain constants for all ranges:
1 Measure the external “10 V” standard on the 
10 V range.
2 Create the gain adjustment for the 10 V range 
using the ratio of the measured and actual 
values.
3 Measure accuracy of the internal reference 
voltage relative to the external standard, and 
store the difference as a reference 
adjustment. (When subsequently invoked, 
auto-calibration uses this stored value to 
re-determine all gain adjustment constants).