3 Adjustment Procedures
40 Keysight 3458A Calibration Manual
Front Terminal Offset Adjustment
This adjustment uses an external 4-terminal short. The multimeter makes offset 
measurements and stores constants for the DCV, DCI, OHM, and OHMF functions. 
These constants compensate for internal offset errors for front terminal 
measurements.
Equipment required: A low-thermal short made of 12 or 14 gauge solid copper 
wire as shown in Figure 3-1 on page 39.
1 Make sure you have performed the steps described previously under 
“Preliminary Adjustment Procedures”.
2 Connect a 4-terminal short across the front panel HI and LO Input terminals 
and the HI and LO W Sense terminals as shown in Figure 3-1 on page 39.
3 After connecting the 4-terminal short, allow 5 minutes for thermal equilibrium.
4 Execute the CAL 0 command. The multimeter automatically performs the front 
terminal offset adjustment and the display shows each of the various steps 
being performed. This adjustment takes about 5 minutes. When the 
adjustment is complete, the multimeter returns to displaying DC voltage 
measurements.
Take precautions to prevent thermal changes near the 4-wire short. You should 
not touch the short after it is installed. If drafts exist, you should cover the input 
terminals/short to minimize the thermal changes.