5 Command Summary
90 Keysight 3458A Calibration Manual
CAL?
Description
Calibration query. Returns a string containing one of four values for the
calibration constant specified; the initial (nominal) value, low limit, high limit, or
actual value of the specified constant. The returned string also contains a
description of the constant. This command is in the full command menu; it is not
in the short command menu.
Syntax
CAL? const_id [,cal_item]
cal_item
Specifies which of the four calibration constant values is to be returned. The
cal_item parameter choices are:
cal_item Description
0 Initial (nominal) value
1
Actual value
[a]
[a] The default for cal_item is the actual value.
3 Upper limit
5 Lower limit