Command Summary 5
Keysight 3458A Calibration Manual 89
For highest accuracy when performing a CAL 0, a four-point short must be
used. Also, CAL 0 must be performed twice, once for the front terminals and
again for the rear terminals. You must manually switch the terminals to be
calibrated using the front panel switch.
It is recommended that the OCOMP command be executed prior to adjusting
with the 10 K source and OCOMP be set to ON. This will account for any
thermals and result in a more accurate adjustment.
Related commands: ACAL, SCAL, SECURE
Example
OUTPUT 722; "CAL 10.0011" !DCV ADJUSTMENT SOURCE = 10.0011 VOLTS
OUTPUT 722;"OCOMP ON"
OUTPUT 722:"CAL 10000.001" !RESISTANCE ADJUSTMENT SOURCE = 10000.001
OHMS