Performance Verification Tests 4
Keysight 3458A Calibration Manual 65
Analog AC voltage performance tests
Required equipment
The following list of equipment is required to test the analog AC performance of
the 3458A.
Stable AC voltage source (Fluke 5700A or equivalent).
Transfer Standard DMM (3458A Opt. 002 within 90 days of Cal.)
Shielded test leads terminated with dual banana plugs (such as
Keysight 11000-60001).
Preliminary steps
1 Make certain that the AC source is properly warmed up.
2 The 3458A requires a 4 hour warm up period. If this has not been done, turn
the instrument ON and allow it to warm up.
3 Execute the ACAL AC function on both the transfer standard DMM and the
UUT. This auto calibration will take approximately 1 minute to complete.
4 If the instrument Self Test has not been run, make certain all inputs are
disconnected and execute the TEST function. The display must read “SELF
TEST PASSED”.
5 Configure the transfer standard DMM as follows:
–ACV
–SETACV SYNC
ACBAND 10,2E6
–RANGE10
–RES .001
–TRIG SGL
LFILTER ON
6 Configure the DMM under test as follows:
–ACV

Table of Contents