8 Frequency Scan Measurement
Save
State Saved Saved in instrument state
Range Off|Flat Color|3D Color
Initial S/W Revision A.10.00
Scan Table On/Off
Enables you to choose whether or not to show Scan Table information in the report.
Key Path
Save, Data, Meas Result, Scan Table
Remote Command
:MMEMory:RESults:SCAN ON|OFF|0|1
:MMEMory:RESults:SCAN?
Example :MMEM:RES:SCAN ON
Preset Not part of Preset, but is reset to ON or by Restore Mode Defaults; survives shutdown.
State Saved Saved in instrument state
Range On|Off
Initial S/W Revision A.10.00
Signal List On/Off
Enables you to choose whether or not to show Signal List information in the report.
Key Path
Save, Data, Meas Result, Signal List
Remote Command
:MMEMory:RESults:SLISt ON|OFF|0|1
:MMEMory:RESults:SLISt?
Example :MMEM:RES:SLIS ON
Preset Not part of Preset, but is reset to ON or by Restore Mode Defaults; survives shutdown.
State Saved Saved in instrument state
Range On|Off
Initial S/W Revision A.10.00
Output Format
Enables user to select the output format of Meas Result. If the Output Format is set to HTML, a .html file
will be saved and a directory that contains the .png file for the screen image will be created. If the Output
Format is set to PDF, a .pdf file will be saved.
Key Path
Save, Data, Meas Result
Remote Command
:MMEMory:RESults:OUTPut HTML|PDF
:MMEMory:RESults:OUTP?