44 Impedance Analyzer
2 Post Repair Procedure
Post Repair Procedures
Table 5 lists the required procedures that must be performed after
the replacement of an assembly. These are the recommended
minimum procedures to ensure that the replacement is
successfully completed.
License file
The license code for the frequency option () and generation ID
(G01) must be installed in E:/license, or the error message will be
displayed.
The license code can be downloaded from the Keysight Software
Manager (KSM) or Keysight Software Licensing (KSL).
Table 5 Post Repair Procedures
Replaced
Assembly or Part
Required Adjustment Correction
Constants (CC)
Verification
A11 Source
Board
Perform the required adjustments
using “A11 Source Board” in Spot
Adjustment of the program.
Troubleshooting Using Diagnostic
Test 32
Perform all of the Performance
Test.
A7 LF Source Perform the required adjustments
using “A7 LF Source” in Spot
Adjustment of the program.
Troubleshooting Using Diagnostic
Test 32
Perform all of the Performance
Test.
A5 TRD Perform the required adjustments
using “A5 TRD” in Spot
Adjustment of the program.
Troubleshooting Using Diagnostic
Test 32
Perform all of the Performance
Test.
A8 VRD Perform the required adjustments
using “A8 VRD” in Spot
Adjustment of the program.
Troubleshooting Using Diagnostic
Test 32
Perform all of the Performance
Test.
A51 DSP Board Perform the required adjustments
using “A51 DSP Board” in Spot
Adjustment of the program.
Troubleshooting Using Diagnostic
Test 32
Inspect the Booting Process
SSD (Solid State
Drive)
Perform the required adjustments
using “SSD” in Spot Adjustment
in the program. Copy from license
file (.lic) under E:/license
Inspect the Booting Process