Keysight E5063A Network Analyzer 19
Installation
Installing Front Handles/Rack Mounting Flanges
Installing Front Handles/Rack Mounting Flanges
The E5063A can be installed on a workbench or in a rack.
This section describes how to install the front handles (Option 1CN) used for
moving or transporting the instrument, and how to install the analyzer in an
equipment rack as part of a measurement system (Option 1CM: without the
handles, Option 1CP: with the handles).
Figure 1-2 Installing front handle/rack-mount kits
The handle kit (Option 1CN) is used for transporting and relocating of the
E5063A.
The rack-mount kit (Option 1CM) includes two flanges (locking side plates) for
mounting the E5063A on a rack (482.6mm/19 inches) conforming to the EIA
Standard.
Option 1CP comes with both, the handle kit and rack-mount kit.
While referring to Figure 1-2 install the handle kit/rack-mount kit by following
these steps.
Step 1. Remove the adhesive-backed trim strip (1) from each side of the outer frame
of the E5063A front panel.
Step 2. Use the provided screws to mount the