Calibration 2
S-Series Oscilloscopes Service Guide 15
 
Running a User Calibration
The user calibration uses signals generated in the oscilloscope to calibrate 
channel scale, offsets, and trigger parameters. 
When to perform a user calibration:
• Standard calibration only—When the oscilloscope’s operating temperature 
(after the 30-minute warm-up period) is more than ±5
 °C different from that of 
the last calibration. Be sure to perform a standard user calibration (do not 
perform the time scale cal)—even if one was recently performed—when 
environmental temperature conditions cause the oscilloscope’s operating 
temperature to change, such as when the oscilloscope is moved to a test rack 
or chamber.
• Standard and time scale cal—When it has been more than 1 year since the last 
time scale calibration, and when you replace the acquisition board.
Performing a user calibration will invalidate the oscilloscope’s Certificate of 
Calibration. If NIST (National Institute of Standards and Technology) traceability is 
required, perform the procedures in Chapter 3 using traceable sources.
Equipment required
Calibration time
It takes approximately 1 hour to run the user calibration on the oscilloscope, 
including the time required to change cables from channel to channel.
The oscilloscope must be warmed up (with the oscilloscope application running) for at least 
30 minutes at ambient temperature before starting the calibration procedure. Failure to allow 
warm up may result in inaccurate calibration.
Equipment Critical specifications Recommended part number
50 Ω BNC Cable 50 Ω characteristic impedance 
BNC cable
54609-61609
10 MHz Signal Source (required for 
time scale calibration)
Frequency accuracy better than 
±41 ppb
53131A with Opt. 012
MSO calibration fixture that ships 
with the oscilloscope (MSO models 
only)
1250-3817
MSO cable that ships with the 
oscilloscope (MSO models only)
54904-61622