1 Remove the air duct from the acquisition board by lifting it straight up.
2 Reposition the oscilloscope so you can access the front panel, and remove the
four M2 screws next to each BNC. When replacing, torque to 3 in-lbs using a
power driver and T-6 bit.
3 Remove the two M3 screws that connect the heat sink to the chassis. Do not
remove the other two screws.