Troubleshooting 4
S-Series Oscilloscopes Service Guide 61
 
Types of Faults
The fact that a fault on one sequencer causes LEDs corresponding to other 
sequencers to illuminate will help you determine which LED is the root of the 
problem. 
When a fault occurs on a given sequencer, it notifies the other three using the 
Fault signal. The other three sequencers also report it as an External Fault. For 
example, if the first sequencer has an over-voltage on channel 1, it will report an 
External Fault with channel 1 as the source. The other three sequencers will report 
it as an External Fault with the source as Channel 4. 
Similarly, if the first sequencer has an under-voltage failure on +12V, it will report a 
Reset Fault on channel 1 and the other three sequencers will report External 
Faults on channel 4. Any fault on one of the sequencers always reports an External 
Fault on channel 4 on the other three sequencers. Look for the fault that is not an 
External Fault on channel 4 as the cause of the shutdown. 
If all four sequencers report an External Fault on channel 4, it is more difficult to 
determine the real source. It could be on any of the Channel 4 sources, an 
over-temperature fault (look for the over-temperature LED to light), or an 
FPGA-issued fault.
Sequence Fault:   Occurs when a supply does not come up in time during a 
power-on sequence.
Reset Fault:   Occurs if the oscilloscope shuts down during the monitor phase 
because a supply went under voltage. This fault is probably due to the supply 
failing to an off state or the load becoming too much for the supply. 
Command Fault:   Occurs when the ON sequence to one of the sequencers goes 
HIGH to initiate a turn-on, but then goes LOW before the turn-on sequencing 
finishes. Similarly, if the ON signal goes HIGH during sequencing off, a Command 
Fault occurs. This kind of fault indicates something is going wrong with the ON 
signal. This problem will most likely happen on the first sequencer. The ON signal 
is driven by the minimum on-time limiter. There could be a problem here as well.
External Fault:   An External Fault can be caused by several failures. If a supply 
goes over-voltage, an external fault is issued. If the over-temperature sensor or 
FPGA issues a fault, it will be displayed as an external fault on channel 4. This 
presents an ambiguous case as an External Fault on channel 4 could occur for 
multiple reasons. To help determine the cause, the over-temperature LED will light 
during an over-temperature fault. However, this LED does not latch to failure. 
Once the oscilloscope cools to below the over-temperature condition, the LED will 
turn off. The main cause of an over-temperature fault is a stopped fan or blocked 
air inlet. Check for these causes and power the oscilloscope back up. If it 
continues to shut down, it could be a bad over-temperature sensor or comparator 
circuit.