84 S-Series Oscilloscopes Service Guide
5 Replacing Assemblies
3 Disconnect the three cables from the rear of the keyboard.
4 Unroute the USB double cable from the cable clips.
5 Disconnect the backlight cable from the front of the keyboard.
6 Pull the USB double cable through the opening to finish removing the
keyboard.