Replacing Assemblies 5
S-Series Oscilloscopes Service Guide 87
Removing and Replacing the Processor Assembly
1 Remove the cover and rear fan assembly as described earlier.
2 Remove the M4 screw connecting the acquisition deck to the processor bracket
on the I/O bracket side near the rear of the instrument.
3 Remove the M4 screw connecting the acquisition deck to the processor bracket
near the power supply.
4 Remove the three M4 screws connecting the acquisition deck to the processor
bracket at the top of the instrument.