Frequency Response Analysis
On G-suffix oscilloscope models (that have a a built-in waveform generator), the 
Frequency Response Analysis (FRA) feature controls the built-in waveform 
generator to sweep a sine wave across a range of frequencies while measuring the 
input to and output from a device under test (DUT). At each frequency, gain (A) 
and phase are measured and plotted on a frequency response Bode chart.
When the frequency response analysis completes, you can move a marker across 
the chart to see the measured gain and phase values at each frequency point. You 
can also adjust the chart's scale and offset settings for the gain and phase plots.
Table 22 Frequency Response Analysis Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Frequency Response 
Analysis (FRA) 
enable/disable
[Analyze] > Features,  Entry knob to select Frequency Response Analysis, push Entry knob to 
enable or disable
Channels probing DUT 
input V and output V
[Analyze] > Setup > Input V
[Analyze] > Setup > Output V
Frequency sweep min 
and max values
[Analyze] > Setup > Min/Max Freq, Entry knob
Waveform generator 
amplitude and 
expected output load
[Analyze] > Setup > Amplitude, Entry knob
[Analyze] > Setup > Output Load (50
Ω, High-Z)
Run the analysis [Analyze] > Run Analysis
Adjust scale and offset 
of Bode plots
[Analyze] > Chart > Gain/Phase Scale/Offset, Entry knob
Autoscale gain and 
phase plots
[Analyze] > Chart > Autoscale
View measured gain 
and phase values
[Analyze] > Move Marker, Entry knob
Save data to USB 
storage device
[Save/Recall] > Save > Format, Frequency Response Analysis data (*.csv)