198
Concepts
Time Gating Concepts
Time gating allows you to see the separate spectrum of radio 1 or radio 2 to
determine the source of the spurious signal, as shown in Figure 15-5
Figure 15-5 Time-Gated Spectrum of Radio 1
Figure 15-6 Time-Gated Spectrum of Radio 2
Time gating lets you define a time window (or time gate) of when a measurement is
performed. This lets you specify the part of a signal that you want to measure, and
exclude or mask other signals that might interfere.