Power On and Check
8
Running Internal Alignments
To meet the instrument performance specifications, the
analyzer must periodically be manually aligned.
1 Connect a BNC cable with the N- BNC adapter between the
CAL OUT and RF IN front panel connectors.
2 After 15- minute warm up, press > Alignment > Align > All
The self alignment takes about 5 minutes.
Check for Instrument Messages
The spectrum analyzer has two categories of instrument
messages: error and warning messages. A error message is
triggered by operation errors, for example, parameter setting
conflicts or data input that is out of the range of a
parameter. An warning message may be triggered by
hardware defects which could result in damage to
instrument.
Here are some tips to check the instrument messages.
1 Check the display to see if any messages display in the
status bar. Press > More > Show Errors to review each
message. Please refer to User’s Guide for detailed system
message descriptions.
2 When you have reviewed and resolved all of the error
messages, press > More > Show Errors > Clear error queue to
delete the messages.
3 Cycle the power to the analyzer and re- check to see if the
Instrument messages are still there.
4 If the error messages cannot be resolved, contact the
Keysight Customer Contact Center for assistance or service.