2 Making Measurements  
Using VZ
LOW
 for Voltage Measurements
36 U1231A/U1232A/U1233A User’s Guide
Using VZ
LOW
 for Voltage Measurements
Set up your multimeter to make a VZ
LOW 
(low input 
impedance) voltage measurement as shown in 
Figure  2- 6. 
Probe the test points and read the display.
Figure 2-5 VZ
LOW
 voltage display
Do not use the VZ
LOW
 function to measure voltages in circuits that 
could be damaged by this function’s low impedance (≈3 kΩ).
Use the VZ
LOW
 function to remove ghost or induced voltages from your 
measurements
Ghost voltages are voltages present on a circuit that should not be 
energized. They are usually caused by capacitive coupling between 
energized wiring and adjacent unused wiring. The VZ
LOW
 function can 
remove ghost voltages from your measurements by dissipating the 
coupling voltage. Use the VZ
LOW
 function to reduce the possibility of false 
readings in areas where the presence of ghost voltages are suspected.
During VZ
LOW
 measurements, the multimeter’s range is locked to 600 V.
The analog bar graph represents the AC+DC voltage value combined.