6 Calibration
A calibration certificate is available as option in paper format.
We recommend to carry out a yearly checking.
7 Accessories
Accessories References Illustrations
1 AA lithium 3.6 V KBL-AA
Safety lock wall mount with padlock KAV-220
Wired extension for class 220 KISTOCK probes
In polyurethane, 5 m length with male and female mini-DIN
connectors
Note: several extensions can be wired in order to obtain up to
25 m cable length
KRB-220
KILOG software: Configuration and data processing software
KILOG software allows to configure, save and process your data
in a very simple way.
Software only: KILOG-3-N
Complete set (software +
1 USB cable): KIC-3-N
Data collector
Collects up to 20 000 000 points from one or several KISTOCK
directly on-site. Results restitution on PC of realised datasets
KNT-320
USB micro-USB cable which allows to plug your KISTOCK
datalogger to your PC
CK-50
Neck size hand strap KDC
Only the accessories supplied with the device must be used.
8 Troubleshooting
Problem Probable cause and possible solution
No value is displayed, only the icons are present.
The display is configured on “OFF”. Configure it on “ON” with the
KILOG software (see page 14).
“hi” or “lo” is displayed
The measurement range is exceeded. There is a problem with the
sensing element.
The display is completely off* and there is no communication
with the computer.
The battery has to be replaced. (see page 15).
The display indicates “- - - -” instead of the measured value. The probe is disconnected. Plug it again to the datalogger.
* For models with display.
16 Troubleshooting