EasyManuals Logo

Kimo KT 220 User Manual

Kimo KT 220
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
3 Technical features
3.1 Technical features of the devices
KT 220 KH 220 KTT 220
Units displayed
°C, °F, °Ctd, °Ftd,%RH,
mV, V, mA, A, bar
1
°C, °F, °Ctd, °Ftd,%RH, lux, fc, mV, V,
mA, A, bar
1
°C, °F
Resolution
0.1°C, 0.1°F,
0.1%RH, 1 mV, 0.001 V,
0.001 mA, 0.1 A, 0.1 bar
0.1°C, 0.1°F, 0.1%RH, 1 lux, 0.1 fc, 1 mV,
0.001 V, 0.001 mA, 0.1 A, 0.1 bar
0.1°C, 0.1°F
External input
Female micro-USB connector
Input for probe 1 universal input
2
1 universal input
2
2 inputs for
thermocouple probes
(K, J, T, N, S)
Internal sensor
Temperature Temperature, humidity, light Temperature
Type of sensor NTC
Temperature: NTC
Humidity: capacitive
Light: photodiode
Thermocouple
Measuring range
Measuring range of the
internal sensor
3
:
From -40 to +70°C
Measuring range of the internal sensor
3
:
Temperature: from -20 to +70°C
Humidity: from 0 to 100 % RH
Light: from 0 to +10 000 lux
K: from -200 to +1300°C
J: from -100 to +750°C
T: from -200 to +400°C
N: from -200 to +1300°C
S: from 0 to 1760°C
Accuracies
4
±0.4°C from -20 to 70°C
±0.8°C below -20°C
Temperature: ±0.4°C from 0 to 50°C
±0.8°C below 0°C or above 50°C
Humidity
5
: ±2%RH from 5 to 95%RH
from 15°C to 25°C
Light: ±10% of reading +10 lux
K, J, T, N: ±0.4°C from 0 to
1300°C
±(0.3% of reading +0.4°C)
below 0°C
S: ±0.6°C
Setpoint alarms 2 setpoint alarms on each channel
Frequency of
measurement
From 1 second to 24 hours
Operating
temperature
From -40 to +70°C From -20 to 70°C From -20 to 70°C
Storage
temperature
From -40 to +85°C
Battery life
4 years
6
European
directives
2011/65/EU RoHS II; 2012/19/EU WEEE; 2014/30/EU EMC; 2014/35/EU
1
Some units are available only with optional probes.
2
Input which allows to plug different compatible probes: see optional probes and cables page 9.
3
Other measuring ranges are available according to the connected probe: see optional probes and cables page 9.
4
All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurement carried out in the same conditions, or carried out with calibration
compensation.
5
Factory calibration uncertainty: ±0.88%RH. Temperature dependence: ±0.04 x (T-20) %RH (if T<15°C or T>25°C)
6
On the basis of 1 measurement each 15 minutes at 25°C.
Technical features 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kimo KT 220 and is the answer not in the manual?

Kimo KT 220 Specifications

General IconGeneral
BrandKimo
ModelKT 220
CategoryData Loggers
LanguageEnglish

Related product manuals