Installation and connection Diaphragm pump N96
16 Translation of Original Operating and Installation Instruction, English, KNF 316836-316838 01/20
7.3 Electrical connection
DANGER
Danger to life from electric shock
àOnly have the pump connected by an autho-
rized specialist.
àOnly have the pump connected if the power
supply is disconnected.
à
When connecting to a power source, observe the applicable standards,
directives, regulations and technical standards.
à
Install in the electrical wiring system a device to disconnect the pump
motor from the electrical supply network (in acc. with EN 60335-1).
à
Protect the pump motors in accordance with EN 60204-1 (overcurrent
protection, overload protection).
à
The motors may only be operated in an SELV circuit.
The control lines of the BLDC motor are only protected to a voltage
of up to 1.5 kV (acc. to HBM ESD rating).
If higher ESD requirements are needed, measures are to be pro-
vided by the owner himself.
Refer to the type plate for the maximum current consumption of the
pump.
à
It is recommended that an additional EMERGENCY-STOP device be
installed.
à
Mount the pumps in such a way that it is not possible to touch the elec-
trically live parts (electrical connection).
The ground potential of the power supply, the interface(s) and the
pump housing are to be at the same potential. Compensating cur-
rents via the motor controller are to be prevented as they may result
in the destruction of the electronics. A sufficient potential equaliza-
tion in accordance with EN 60479-1 is to be dimensioned.
Fastening the connection cables
à
Fasten the connection cables so that
the cables do not come into contact with movable or hot parts.
the cables cannot be worn or damaged on sharp corners or edges
no tensile and pressure forces are exerted on the connection point
of the cables (strain relief)