Electrical Requirements
KODAK 9000 3D Extraoral Imaging System_Installation Guide (SM711)_Ed 01 4–5
A single-phase alternating current power supply is required and the electrical installation
specifications should be as follows:
Table 4–2 Electrical Installation Specifications
Supply Voltage
230V/240V
100V/110V/
130V
Frequency 50/60 Hz 50/60 Hz
Electrical supply 6 kW 6 kW
Line current
required
16A 20A
Cable cross-section
according to length
For 30 m:
2.5 mm2
For 10 m:
2.5 mm2
Max. line impedance 0.5 0.12
Differential circuit
breaker (2)
30 mA 30 mA
Maximum current to
trip circuit breaker
16A 20A Protect the power supply line with a
differential circuit breaker that trips at
maximum current.
Specifications of 2
red color actuator
emergency stop
push-button
(3) and (4)
250V 6A
UL listed
250V 6A
UL listed
Install these stop push-buttons to
simultaneously, switch off the
current to the active conductors of
the radiological installation and
exclude any other electric
equipment.
Locate (3) inside the x-ray room,
near the Unit, for the patient to
quickly cut the power supply if
necessary.
Locate (4) outside the x-ray room,
near the x-ray remote control for
the operator to quickly cut the
power supply if necessary.
Maintain them in OFF (open)
position until a deliberate action is
performed.
Specifications of the
warning lamps
(5) and (7)
60 W 60 W
Locate the red warning lamp (5)
outside the x-ray room to indicate
the Unit is active (1 lamp at each
access point).
Locate the green warning lamp (7)
outside the x-ray room to indicate
the x-ray emission.

Other manuals for Kodak 9000