X-Ray Room Requirement
KODAK 9000 3D Extraoral Imaging System_Installation Guide (SM711)_Ed 01 4–7
X-Ray Room Requirement
IMPORTANT
Use an appropriate wall fixing system suitable for the type of wall
construction. See the examples below.
The following illustrations provide examples of wall types and fixations.
IMPORTANT
Install the Unit where a minimum amount of space is provided to allow
easy access for the patient or the maintenance technician (see Figure 4-2).
Table 4–3 X-Ray Room Requirements
Room Components Minimum Requirement Comments
Width of the door 75 cm (30")
Height of the ceiling 240 cm (95") If needed, a shorter column is possible
by special order.
Strength of the wall Withstanding an extraction force of 150 kPa
at each point of attachment.
It is the responsibility of the installer to
choose an appropriate fixing system that
withstands the extraction force.
Load-bearing capacity of
the floor
500 kg/m2 The floor must be solid and flat.
Required space for the
KODAK 9000 3D Unit
1500 (L) x 2000 (D) x 2400 (H) mm