X-Ray Room Requirement
4–8 SITE PREPARATION BEFORE INSTALLATION
Figure 4–2 Minimum X-Ray Room Space Configuration
WARNING
You MUST prepare the appropriate electrical requirements and
configurations of the x-ray room before installing the Unit. You MUST
locate separately the following low voltage and high voltage connections:
- Low Voltage: (12), (9), (10)
- High Voltage: (6), (5), (7bis)
3 Red color actuator emergency stop push-button 9 X-ray remote control
4 Red color actuator emergency stop push-button 10 Door safety switch
5 Red warning lamp, power ON indicator 11 Mains outlet (for electrical tools)
6 Unit mains connecting terminal 12 Ethernet outlet (RJ45/1)
7 Green warning lamp, x-ray emission indicator 13 Local Area Network (LAN), (RJ45/2)
7bis X-Ray warning lamp connecting terminal