Fitting the Covers
KODAK 9000 3D Extraoral Imaging System_Installation Guide (SM711)_Ed 01 6–
Fitting the Covers
WARNING
Before fitting the covers, you MUST switch off the Unit.
Before fitting the covers check that:
All the screws in the wall and the floor have been tightened
All the screws attaching the head and the fixed arm to the column have been tightened
To fit the covers, follow these steps:
1. Fit the upper cover (A) on the column. Fit the cover on the wall-bracket (B). Fit the cover
on the column connecting terminal (C) and tighten the 4 screws.
2. Fit the side covers (D) on the column. Position and attach the ON/OFF button to the cover
securely with the 2 screws. Fit the cover (E) to the inside of the column. Tighten the
screws.
3. Fit the 2 covers (F) and (G) under the head and tighten the screws.
4. Fit the Head cover (H) of the Unit to the head chassis and tighten the 4 screws. You can
switch on the Unit now.
H
A
B
D
C
F
G
E
3 mm
2 x M5x12
4 x
M5x50
2 x
M5x25
3 x
M5x12
4 x
M5x12