Fitting the Covers
KODAK 9000 3D Extraoral Imaging System_Installation Guide (SM711)_Ed 01 6–
Fitting the Covers
WARNING
Before fitting the covers, you MUST switch off the Unit.
Before fitting the covers check that:
• All the screws in the wall and the floor have been tightened
• All the screws attaching the head and the fixed arm to the column have been tightened
To fit the covers, follow these steps:
1. Fit the upper cover (A) on the column. Fit the cover on the wall-bracket (B). Fit the cover
on the column connecting terminal (C) and tighten the 4 screws.
2. Fit the side covers (D) on the column. Position and attach the ON/OFF button to the cover
securely with the 2 screws. Fit the cover (E) to the inside of the column. Tighten the
screws.
3. Fit the 2 covers (F) and (G) under the head and tighten the screws.
4. Fit the Head cover (H) of the Unit to the head chassis and tighten the 4 screws. You can
switch on the Unit now.