EasyManua.ls Logo

Kodak DirectView CR 850 - Using the Diagnostics; System Service Diagnostic Screen

Kodak DirectView CR 850
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DIAGRAMS
10 08NOV06 DI4825-1
Section 3: Functional
Power Distribution
K1
RELAY
MP4 POWER SUPPLY
V1 24V @ 8.5A
V2/V3 12V @ 10A / 24V @ 2A
V1
24V
BLUE
BROWN
GREEN/YELLOW
V2
V3
12V
-24V
+
-
+
-
+
-
POWER SUPPLY ASSEMBLY
Neutral
Line
Gnd
F1
8A
GND
V3 USED AS -24V
FOR GALVO BD.
4
3
2
1
RED
RED
RED
BLK
BLK
BLK
BLK
PS1
WHITE
BLACK
GREEN
YELLOW
YELLOW
BLACK
BLACK
BLACK
PURPLE
RED
RED
BLK
BLK
BLK
BLK
BLK
TB2
56
432 1
PURPLE
AC
SOURCE
90-264 VAC
50-/60 HZ.
{
WORLDWIDE
TRANSFORMER
W/ ISOLATION
1000VA UPS
120VAC 50/60HZ
AC
AC
RS232
AC
AC
N/C
T1
N/C
UPS1
L
N
G
P
.
C
.
Comm 3
P. C .
AC Power
1
2
J1
YELLOW
BLACK
1
2
P1
CF1
FAN
To P3 Touch Screen Display
To P2 GALVO BD.
To P5 MSC BD.
E2
To P5 MSC BD.
To P5 MCPU BD.
To PMT / DAS BD.
To P1 SLOW SCAN CONTROLLER BD.
To
J1
Erase Assembly
To P10 MSC Bd.
To P3
MSC Bd.
AC