Kodak DirectView CR 850 - User Manual

Kodak DirectView CR 850
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
© EASTMAN KODAK COMPANY, 2005 HEALTH GROUP
Confidential
Restricted
Information
{Diagnostics}{Production}{Health Group}{ExternalAndInternal}
Publication No. DG4825-1
10DEC05
Supersedes DG4825-1
20JUL04
DIAGNOSTICS
for the
Kodak DirectView CR 825/850 SYSTEMS
Service Codes: 5634, 4825
Important
Qualified service personnel must repair this equipment.
H177_0500GC

Table of Contents

Other manuals for Kodak DirectView CR 850

Questions and Answers

  • A
    Andrew KellyAug 9, 2025
    What to do if the TOUCH SCREEN of Kodak DirectView CR 850 Medical Equipment is dark?
    • J
      Jill WhiteAug 9, 2025
      If the TOUCH SCREEN is dark, check the power to the host computer and POWER SUPPLY PS1.
  • A
    ashley10Aug 30, 2025
    What to do for application software errors in Kodak DirectView CR 850?
    • G
      Greg WilsonAug 30, 2025
      To resolve application software errors in Kodak Medical Equipment, install the application software again.
  • M
    michael88Nov 18, 2025
    What causes initialization errors in Kodak Medical Equipment?
    • J
      James RamosNov 18, 2025
      Initialization errors in Kodak Medical Equipment can be due to several reasons. If the TOUCH SCREEN displays a POST Error, see the 'Troubleshooting Section' of the Owner’s Manual for the IBM COMPUTER. Listen to the beeps when the computer energizes. Initialization errors may also occur due to failure of the video ROM, the 'Primary DMA Register' test, the 'Secondary DMA Register' test, the 'Primary interrupt mask register' test, the 'Secondary interrupt mask register' test, the 'Interrupt vector loading' test, the test for the KEYBOARD CONTROLLER, power failure of the CMOS and checksum failure, failure of the CMOS, failure of the initialization of the screen, failure of the 'Screen Memory' test, or failure of the 'Screen Retrace' test. In some cases, installing a new CPU may resolve the iss...

Summary

Using the Diagnostics

System Service Diagnostic Screen

Provides access and overview of the diagnostic system interface.

Component Tests

Allows actuation and status check of system components.

Diagnostic Tests

Includes Light Test and Slow Scan Current/Velocity Tests.

Component Tests and Plots

Covers Laser Power, Galvo, PMT tests, and Galvo Plot.

System Self Test

Completes diagnostic tests on various system boards.

Diagnostic Tests

Individual Component Control

Tests individual components by actuating them and checking sensor status.

Component Control: Step

Guides through loading, scanning, erasing, and ejecting cassette and plate.

Slow Scan Current and Velocity Tests

Drives SLOW SCAN AY to limits and checks current and velocity.

Laser Power Test

Checks and sets the power of the LASER DIODE.

Galvo Test Procedure

Tests GALVO movement and LASER DIODE response.

MCPU Self Test

Performs self-tests on the MCPU board.

Error Codes and Troubleshooting

System Troubleshooting and Checkout

Initialization Errors

Errors occurring if the system energizes but does not display the main menu.

“Power-On Self-Test” (POST) Errors

Describes beep codes and their corresponding malfunctions during POST.

BOARDS Checkout

Checkout procedures for various system boards.

Test Points and LEDs

MOTION SYSTEM CONTROL BOARD A1 Test Points and LEDs

Diagrams and lists test points and LEDs on the board.

MASTER CENTRAL PROCESSING UNIT BOARD A2 Test Points and LEDs

Diagrams and lists test points and LEDs on the board.

Kodak DirectView CR 850 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 850
CategoryMedical Equipment
LanguageEnglish