What to do if the TOUCH SCREEN of Kodak DirectView CR 850 Medical Equipment is dark?
- JJill WhiteAug 9, 2025
If the TOUCH SCREEN is dark, check the power to the host computer and POWER SUPPLY PS1.




What to do if the TOUCH SCREEN of Kodak DirectView CR 850 Medical Equipment is dark?
If the TOUCH SCREEN is dark, check the power to the host computer and POWER SUPPLY PS1.
What to do for application software errors in Kodak DirectView CR 850?
To resolve application software errors in Kodak Medical Equipment, install the application software again.
What causes initialization errors in Kodak Medical Equipment?
Initialization errors in Kodak Medical Equipment can be due to several reasons. If the TOUCH SCREEN displays a POST Error, see the 'Troubleshooting Section' of the Owner’s Manual for the IBM COMPUTER. Listen to the beeps when the computer energizes. Initialization errors may also occur due to failure of the video ROM, the 'Primary DMA Register' test, the 'Secondary DMA Register' test, the 'Primary interrupt mask register' test, the 'Secondary interrupt mask register' test, the 'Interrupt vector loading' test, the test for the KEYBOARD CONTROLLER, power failure of the CMOS and checksum failure, failure of the CMOS, failure of the initialization of the screen, failure of the 'Screen Memory' test, or failure of the 'Screen Retrace' test. In some cases, installing a new CPU may resolve the iss...
Provides access and overview of the diagnostic system interface.
Allows actuation and status check of system components.
Includes Light Test and Slow Scan Current/Velocity Tests.
Covers Laser Power, Galvo, PMT tests, and Galvo Plot.
Completes diagnostic tests on various system boards.
Tests individual components by actuating them and checking sensor status.
Guides through loading, scanning, erasing, and ejecting cassette and plate.
Drives SLOW SCAN AY to limits and checks current and velocity.
Checks and sets the power of the LASER DIODE.
Tests GALVO movement and LASER DIODE response.
Performs self-tests on the MCPU board.
Errors occurring if the system energizes but does not display the main menu.
Describes beep codes and their corresponding malfunctions during POST.
Checkout procedures for various system boards.
Diagrams and lists test points and LEDs on the board.
Diagrams and lists test points and LEDs on the board.