DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
97 of 180
16006 - Calibration Error: Did not acquire image
16010 - Possible Light Leak
16011 - Imaging System Error occurred during Initialization
16012 - PMT/DAS BOARD A5: Calculation Error occurred during
Initialization
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action: Do the checkout for the DIGITIZER BOARD A3.
Cause 3: The GALVO BOARD A4 malfunctioned.
Action: Do the checkout for the GALVO BOARD A4.
Description: Large quantities of light occurred during the “AUTO-ZERO.
Cause: The DOOR is open, a PANEL is loose, or a light leakage occurred in the
CASSETTE HANDLING SYSTEM.
Action: 1. Do the Light Test to locate the source of the light leakage.
2. Repair the light leakage.
Cause 1: The PMT/DAS BOARD A5 malfunctioned.
Action: 1. Check the CABLE between the PMT BOARD and the DAS BOARD.
2. Do:
checkout for the PMT/DAS BOARD A5
MCPU Self Test
Description: The value for one or more of the following parameters did not match the
specification when the CR 825/850 SYSTEM initialized: “Fine or Coarse
Offset”, or “Fine or Coarse Ratio.

Table of Contents

Other manuals for Kodak DirectView CR 850

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 850 and is the answer not in the manual?

Kodak DirectView CR 850 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 850
CategoryMedical Equipment
LanguageEnglish

Related product manuals