DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
73 of 180
12101 - SLOW SCAN a velocity malfunction during scan
12102 - SLOW SCAN stopped during scan
12103 - SLOW SCAN stopped during scan
12104 - SLOW SCAN: Timeout while moving to SENSOR S9
Description: The SLOW SCAN AY exceeded the speed limit when moving the SCAN
PLATE.
Cause 1: The speed of the SLOW SCAN AY exceeded the limit. A band might appear on
the image.
Action: Do the Slow Scan Current and Velocity Tests.
Cause 2: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: 1. Check for an obstruction.
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: 1. Check for an obstruction.
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The SLOW SCAN AY did not reach the EXTRACTION BAR SENSOR
S9.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.