EasyManua.ls Logo

Kodak DirectView CR 850 - 12109 - SLOW SCAN Stopped During Scan; 12107 - SLOW SCAN: +24 V DC Malfunction During Scan; 12108 - SLOW SCAN Stopped During Scan; 12110 - SLOW SCAN DID Not Communicate with the MOTION SYSTEM CONTROL

Kodak DirectView CR 850
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
74 of 180
12107 - SLOW SCAN: +24 V DC malfunction during scan
12108 - SLOW SCAN stopped during scan
12109 - SLOW SCAN stopped during scan
12110 - SLOW SCAN did not communicate with the MOTION SYSTEM
CONTROL
12111 - SLOW SCAN did not move
Cause 1: The POWER SUPPLY PS1 malfunctioned.
Action: Do the checkout for the POWER SUPPLY PS1.
Cause 2: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The SLOW SCAN AY did not move because of a communication error
with the SLOW SCAN CONTROLLER BOARD A6.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.

Table of Contents

Other manuals for Kodak DirectView CR 850

Related product manuals